7. Introduction to Cryo-EM#
Cryogenic electron microscopy (cryo-EM) determines the structure of biological specimens that are preserved in a thin layer of vitreous (amorphous) ice at cryogenic temperature (Section 6.3). It has become the dominant method for determining the three-dimensional structures of proteins and macromolecular complexes and earned the 2017 Nobel Prize in Chemistry for Jacques Dubochet, Joachim Frank and Richard Henderson. The electron beam passes through the thin, hydrated specimen, which is preserved in a state close to its native conformation in solution. In contrast to X-ray crystallography, no crystals are needed.
Two families of methods use such specimens. Single-particle analysis (SPA) averages tens of thousands to millions of images of identical, randomly oriented copies of a biological molecule to reconstruct a single three-dimensional density map, often to near-atomic resolution. Electron tomography (cryo-ET) records a series of images of one specimen at different tilt angles and combines them into a three-dimensional volume. It gives a unique view of structurally heterogeneous objects, such as sections of cells, organelles or protein assemblies as they occur in situ in native cells.
Both methods face the same problem that biological specimens scatter electrons weakly and tolerate only a small electron dose, which makes that the resulting images are extremely noisy. This chapter introduces this problem, the principle of averaging by which the signal can be recovered, and the two strategies by which projections of a specimen are recorded. The last section shows how the following chapters build on it.
7.1. Contrast and dose limits in cryo-EM#
Biological macromolecules consist of light elements (hydrogen, carbon, nitrogen and oxygen) and scatter electrons weakly (Section 5.1). Two consequences determine how cryo-EM images are recorded.
Weak contrast. A thin, unstained specimen removes only a small fraction of the electrons from the beam (a few percent for a small protein), so that its amplitude contrast is almost zero. The structural information is carried by the phase of the electron wave. It is converted into intensity by the defocus of the objective lens, and only at the spatial frequencies that are selected by the contrast transfer function. These principles are explored in (Section 10).
Radiation damage. Electron interactions with the sample also deposit energy in the specimen, which breaks chemical bonds and gradually destroys the structure (Section 5.4). The cryogenic temperature at which the specimen is held slows this process but does not prevent it. The effect on the image depends on the spatial frequency. The signal that the specimen contributes at spatial frequency \(k\) fades with the fluence \(D\) (also called dose, in e⁻/Ų) it has already received approximately as \(\exp(-D/D_\mathrm{c}(k))\). The critical fluence \(D_\mathrm{c}(k)\) is the fluence after which it has fallen to \(1/e\), and it decreases with increasing \(k\), so that fine detail (e.g. side chain density) fades at a lower dose than secondary or tertiary structure. For an average protein imaged at 300 kV, \(D_\mathrm{c}\) ≈ 4 e⁻/Ų at 3 Å and 11 e⁻/Ų at 8 Å. Electrons that arrive late therefore add noise but little signal at high resolution. For an image that sums all electrons up to a fluence \(D\), the signal-to-noise ratio at a given resolution is largest at about 2.5 \(D_\mathrm{c}\). Beyond this exposure, further electrons lower it again, though only slowly.
Because of the radiation sensitivity, cryo-EM imaging has to operate under a strict dose budget. In low-dose imaging, the area of interest is never illuminated before the recording. The specimen is searched and the microscope is focused on neighbouring areas, and the final image is recorded as a movie of many short frames (Section 8.3.1). In single-particle analysis, the total dose is typically 30–60 e⁻/Ų. This is more than the optimal exposure for the finest detail, but the signal at lower resolution, on which the alignment of the particles depends, continues to increase, and the frames are weighted according to the accumulated dose during processing (see Section 11.6.1). In tomography, the total dose is of the order of 100 e⁻/Ų and has to be shared across all the individual images of the tilt series (Section 12.4.1). Weak contrast and a limited dose together mean that a single image is dominated by noise.
7.2. Signal, noise, and the need for averaging#
Electrons are discrete quanta. When on average \(\bar{n}\) electrons arrive in a pixel, their number follows a Poisson distribution with variance \(\bar{n}\), so that the irreducible shot noise of the beam has a standard deviation of \(\sqrt{\bar{n}}\). In cryo-EM, a typical dose of 50 e⁻/Ų spread over a micrograph with 1 Å pixels delivers about 50 electrons per pixel, with a shot noise of about \(\sqrt{50} \approx 7\) electrons per pixel. The signal of interest, the contrast modulation produced by a single protein particle embedded in ice, is only a small part of the mean count. Shot noise is only the lower limit of the noise in an image. Structural noise, the irreproducible component of the object before the image is formed (for example the ice and the partially ordered solvent around the particles), is of the same order of magnitude as the reproducible signal. Inelastically scattered electrons add further background as we will discuss in Section 10).
The signal-to-noise ratio (SNR) is defined here as the ratio of the signal power to the noise power,
with the variance \(\sigma_\mathrm{s}^2\) of the signal and the variance \(\sigma_\mathrm{n}^2\) of the noise. For the image of a single particle it is typically 0.01–0.1, so that the protein is often almost invisible in a single raw image. The solution to improving this situation is coherent averaging. If \(N\) images of identical objects are aligned and averaged, the signal, which is the same in each image, is unchanged, whereas the variance of the independent noise is reduced by the factor \(N\). The SNR of the average is therefore
The ratio of the amplitudes, \(\sqrt{\mathrm{SNR}}\), increases in proportion to \(\sqrt{N}\). To improve it by a factor of 10, one needs \(N = 100\) images, and a factor of 100 requires \(N = 10^4\) images. This holds for aligned images of the same object in the same view, as in the example below and in the average of a class of similar particle views. It is slightly different for three-dimensional reconstruction, because the particles are projections from many directions but also here averaging significantly improves the SNR as we will discuss in Section 12.1.4). The interactive below demonstrates how shot noise degrades images at low dose and how averaging recovers the underlying signal. In this example, the images are perfectly aligned. In practice, the orientations of the particles are unknown, and the gain is only realised if they are determined from the images themselves.
Interactive element
Click the rocket symbol at the top of the page and select Live Code, or expand the Show code toggle and click ▶ to run the cell. The dose slider sets the number of electrons per Ų per image. The single images show almost no recognisable structure, whereas the image quality of the average improves with the number of images that are averaged.
7.3. Single particles and tilt series#
A TEM image of a thin specimen is to good approximation a two-dimensional projection of its three-dimensional Coulomb potential (the projection approximation, Section 10.1). A single projection does not determine the structure, because a volume has more unknowns than a projection has pixels. Instead, projections from many directions are required that need to be combined to reconstruct the three-dimensional object. If the same molecule were imaged repeatedly from different directions, the accumulated dose would destroy the high-resolution information before enough projections were collected. In single-particle analysis, each projection is therefore taken from a different molecule, i.e. many identical copies of the molecule in random orientations are imaged once each, and the unknown orientations are determined computationally. In electron tomography, the total dose is divided over a tilt series of a single specimen, which provides known projection angles and the three-dimensional context of a unique object but a lower signal per projection. For repeated copies of a structure within a tomogram, subtomogram averaging combines the advantages of both approaches.
7.4. From the specimen to the structure#
The chapters of this part follow the path from the specimen to the structure, and each of them addresses a problem that has been introduced here. The preparation of vitrified specimens has been described in Section 6.3.
Section 8 describes the microscope itself. It shows how the electron beam is produced and focused onto the specimen, how the lenses form and magnify the image, and how a detector records it.
Section 9 introduces the Fourier transform, which we use to describe images, their resolution and the way the microscope transfers them.
Section 10 explains how a weakly scattering specimen produces contrast by changing the phase of the electron wave, and how defocus turns this phase shift into intensity. It introduces the contrast transfer function which describes how strongly each spatial frequency is transferred into the image, and shows how to correct for it or modify it with a phase plate…
Section 11 shows how thousands of noisy images of identical particles in unknown orientations are aligned, sorted into classes and averaged into a three-dimensional map.
Section 12 deals with the other strategy, in which a single specimen is tilted and imaged many times. It describes how the tilt series is reconstructed, why a wedge of missing data remains and how averaging repeated structures within a tomogram can mitigate its effect and improve the resolution.