8. The Transmission Electron Microscope#
In a transmission electron microscope (TEM), the image is formed from electrons that pass through an electron-transparent specimen (Section 4.1 compares the imaging modes). In conventional SEM (Section 4), contrast usually comes from secondary or backscattered electrons (Section 5.2 and Section 5.3) generated near the specimen surface, although sectioning can expose internal structures for SEM imaging and backscattered electrons can provide compositional contrast from a finite depth below that surface, or, in case of thin tissue sections, an entire section depth. The transmitted electrons carry information on the internal structure of the specimen over its whole thickness, which makes the TEM the instrument of cryo-EM and cryo-ET.
This chapter describes how the microscope is built and how it meets the demands of weakly scattering, radiation-sensitive specimens introduced there: how electrons are produced, accelerated and focused in the TEM column, how the ray paths through the objective and projector lenses determine the magnification and the image planes, how the image is recorded by a detector that retains as much signal as possible, and which factors limit the resolution. The formation of image contrast from the interaction of the electron wave with the specimen, which relies on the wave nature of the electrons and leads to the contrast transfer function, is described in Section 10, after a summary of Fourier analysis in Section 9.
8.1. The TEM column#
A transmission electron microscope is organized as a tall vertical column kept under high vacuum (\(\sim\) 10-7–10-9 mbar). The vacuum is essential for two reasons: electrons scatter strongly from gas molecules, and many components (especially the electron gun) are degraded by oxidation. The column can be divided into the illumination system, which comprises the electron gun and the condenser lenses and delivers the beam to the specimen; the specimen region, in which the specimen is held by the stage within the objective lens; and the imaging system, which consists of the objective lens, the intermediate and projector lenses and the detector, and which forms and magnifies the image. Fig. 8.1 shows the arrangement of these optical elements along the TEM column.
Fig. 8.1 Schematic of the TEM column. The electrons that leave the gun are focused by the double condenser (C1 and C2, with the C2 aperture) onto the specimen. The objective lens forms the first magnified image, in which the objective aperture is located in the back focal plane, and the intermediate and projector lenses magnify the image further onto the detector.#
8.1.1. Electron gun#
Electrons are produced at the top of the column by an electron gun (different types of electron sources are discussed in Section 3.7). Modern biological TEMs typically use a field emission gun (FEG; see Section 3.7.2 and Fig. 3.5), in which electrons are extracted from a sharp tungsten or ZrO₂-coated tip by a strong electric field. FEGs produce a highly coherent, bright beam (Section 3.7.4) with a small energy spread (\(\Delta E \approx 0.3\)–\(0.8\) eV), which is important for avoiding chromatic aberrations. The electrons are then accelerated through a voltage of typically 80–300 kV. The accelerating voltage determines the electron wavelength via the relativistic de Broglie relation:
At 300 kV, \(\lambda \approx 1.97\) pm; this is roughly \(10^5\) times shorter than visible light, which is why electrons can in principle resolve features below 1 Å. Table 8.1 shows that the relativistic correction cannot be neglected. At 300 kV the electrons travel at about 78 % of the speed of light, and the non-relativistic formula overestimates the wavelength by 14 %.
Accelerating voltage (kV) |
\(\lambda\) relativistic (pm) |
\(\lambda\) non-relativistic (pm) |
\(v/c\) |
|---|---|---|---|
80 |
4.18 |
4.34 |
0.50 |
100 |
3.70 |
3.88 |
0.55 |
120 |
3.35 |
3.54 |
0.59 |
200 |
2.51 |
2.74 |
0.70 |
300 |
1.97 |
2.24 |
0.78 |
8.1.2. Illumination system#
Below the gun, a series of condenser lenses (C1 and C2, and in many microscopes a third condenser, C3, or a condenser mini-lens; see Section 3.8) shape the beam before it reaches the specimen. The first condenser lens (C1) controls the effective demagnification of the electron source (the gun crossover) and consequently influences the beam current and effective source size. This is the spot size setting of the microscope. A higher spot size number corresponds to a stronger C1, a smaller image of the source and a lower beam current. The second condenser lens (C2) controls the spreading of the beam and affects the illuminated area and the convergence angle at the specimen. The condenser aperture (C2 aperture) positioned below C2 restricts the transmitted electron beam and helps to control the beam current and illumination geometry. Depending on the condenser lens settings, it can also limit the angular spread of the incident electrons and improve spatial coherence. Fig. 8.2 illustrates how the C1 setting, the spot size, and the C2 aperture together determine the beam current.
Fig. 8.2 Effect of the spot size. A strong C1 lens (high spot size number) forms a small, strongly demagnified image of the source and leads to a small probe at the specimen. The beam is then strongly divergent behind the C1 crossover, so that a larger fraction of the electrons is excluded by the C2 aperture (blue areas) and the current is lower. A weak C1 lens (low spot size number) produces a larger crossover and probe, and more of the electrons pass the aperture.#
In microscopes equipped with a third condenser lens (C3), the additional lens provides an independent control over the beam diameter and convergence at the specimen. The illuminated area can then be changed while the illumination remains nearly parallel, whereas with two condenser lenses a change of the beam diameter is coupled to a change of the convergence. For cryo-EM, this is particularly useful because the beam can be matched to the field of view without changing the illumination angle or the spatial coherence (Section 3.7.4). The ray simulation in the next section includes the third condenser as an option.
A further refinement concerns the edge of the illuminated area. If the beam is limited by a defocused condenser aperture, Fresnel diffraction produces fringes at its edge, and the beam has to be made larger than the detector to keep them out of the image. In the fringe-free illumination (FFI) that is used in modern cryo-EM, the condenser lenses are arranged such that the C2 aperture is imaged onto the specimen plane, whereas the source is imaged into a plane that determines the illumination angle. This is analogous to Köhler illumination in light microscopy, in which the field-limiting aperture and the light source are associated with different sets of conjugate planes. The illuminated area can then match the recorded field closely, which avoids exposing the surrounding specimen to the beam and allows more exposures to be recorded within a single grid hole.
The electrons that are transmitted and scattered by the specimen are subsequently combined by the objective and the imaging system into the bright-field image that is recorded by the detector.
8.1.3. Objective lens#
The objective lens is the most critical optical element in the TEM. It is a strong magnetic lens whose pole piece surrounds the specimen, so that the specimen is located within the field of the lens. The field above the specimen acts as the final condenser lens, which focuses the beam onto the specimen, and the field below the specimen forms the first magnified image. The focal length of the objective is short, of the order of 1–3 mm, and the first image is magnified by a factor of about 30–50. All further lenses only enlarge this image. The aberrations of the objective lens, in particular the spherical aberration \(C_s\) and the chromatic aberration \(C_c\) (Section 2.1.2), which are of the order of the focal length itself, therefore determine the resolution of the microscope, and minimising them is the central engineering challenge of TEM design.
The simulation below shows the ray paths for a thin lens, in which the specimen is drawn in front of the lens for clarity and the scattering angle is exaggerated. Rays that leave the specimen in the same direction are focused in the same point of the back focal plane, at a distance \(f_\theta\) from the axis for a scattering angle \(\theta\), independent of the position in the specimen from which they originate. The unscattered beam is therefore focused on the axis, and the scattered beams are distributed around it. The intensity in this plane is the diffraction pattern, which is the Fourier transform of the electron wave behind the specimen (Section 9). Rays that start from the same point of the specimen are brought together again in the image plane, where they form the magnified image. The projector system can display either plane on the detector by changing the strength of the intermediate lens, which is the difference between the imaging mode and the diffraction mode.
Ray diagram
Rays from two points of the specimen (dark and light) are shown. The two sliders for the position move these points along the specimen, and thus change the positions of the rays on the specimen. The unscattered rays (blue) and the rays that are scattered by \(\pm\theta\) (orange) from one point leave the lens in different directions, but they meet again in the same point of the image plane, which lies at three times the distance from the axis and on the opposite side. The scattering angle sets where the scattered rays cross the back focal plane. If the radius of the objective aperture is smaller than \(f_\theta\), the scattered rays are blocked and only the unscattered rays reach the image, which shows how the aperture removes the scattered electrons. The last slider changes the focal length \(f\) of the lens, in per cent of the value \(f_0\) for which the image plane is exactly in focus, so that a value of zero is the focused image. A change of the focal length corresponds to a change of the defocus: the rays of one point then no longer meet in the image plane, and the image of a point becomes a blurred spot.
Interactive element
Click the rocket symbol at the top of the page and select Live Code, or expand the Show code toggle and click ▶ to run the cell. Move the two points of the specimen, change the scattering angle, the radius of the objective aperture and the focal length of the lens, and observe the ray paths.
Because the back focal plane contains the scattered electrons sorted by their scattering angle, it is the position at which the image formation can be influenced. An objective aperture with radius \(r\) in this plane blocks all electrons that are scattered by more than \(\theta_\mathrm{max} = r/f\). As the scattering angle corresponds to the spatial frequency \(k = \theta/\lambda\) of the specimen, the aperture limits the resolution of the image to \(d = \lambda/\theta_\mathrm{max}\) and removes electrons that were scattered to high angles, which increases the amplitude contrast (Section 10). The aperture is therefore a trade-off between contrast and resolution. In high-resolution cryo-EM, it is often not inserted, in order to retain high-resolution information, and the information limit is determined by the envelope of the CTF. The back focal plane is also the position of the phase plates that are described in Section 10.6.
The defocus of the image is set by changing the current of the objective lens, and the specimen is brought to the height at which this lens focuses on it (the eucentric height) by moving the stage. Because the lens field is never perfectly rotationally symmetric, a pair of stigmators is used to correct the remaining astigmatism, which would otherwise make the image depend on the direction of the spatial frequency (Section 10.5).
Astigmatism makes the defocus depend on the direction in the image plane: the rings of the contrast transfer function (Section 10.4), and thus the Thon rings in the power spectrum of an image, become ellipses (Section 10.5). The two stigmators are electromagnetic quadrupoles (or octupoles) below the objective lens. Each of them generates an astigmatism of its own, along two orthogonal axes, which is adjusted to compensate the astigmatism of the lens. In practice, the microscope operator observes the Fourier transform of a live image of an amorphous region (e.g., the specimen support or the ice) and adjusts the two stigmators until the rings are circular.
The simulation below demonstrates this. The lens has an astigmatism that is unknown to the operator, which is compensated with the two stigmator sliders. The image contains an amorphous background, as for a specimen on a carbon film, so that the Thon rings appear in the power spectrum.
Interactive element
Click the rocket symbol at the top of the page and select Live Code, or expand the Show code toggle and click ▶ to run the cell. The two stigmator sliders add the two components of the astigmatism. Adjust them until the rings of the 2D CTF and of the power spectrum are circular. The button selects another astigmatism setting of the lens. The checkbox switches the amorphous background of the specimen support on and off, and the noise slider sets the noise of the detection; without both, the image of the duck is noise-free, but the Thon rings are then no longer visible in the power spectrum. The defocus slider changes the mean defocus and thus the ring spacing, and the astigmatism is more difficult to judge at a small defocus because there are fewer rings.
8.2. Ray diagrams and paraxial optics#
Electron trajectories through the TEM column can be understood using paraxial ray tracing: the approximation that all rays travel close to the optical axis and at small angles, so that lenses can be represented as thin lenses characterised by a single focal length \(f\). Under this approximation, a thin lens transforms ray slope \(u = dr/dz\) according to:
where \(r\) is the ray height at the lens plane and \(f\) is the focal length. Between lenses, rays travel in straight lines: \(r' = r + u \cdot \Delta z\).
This section is an interlude between the lenses that illuminate and image the specimen and the projector system and the detector, which are described afterwards. The simulation below traces rays through a simplified TEM illumination system with two condenser lenses and an objective. You can adjust focal lengths and aperture sizes to explore how the beam is shaped before reaching the specimen. The C2 aperture below the second condenser limits the beam and thus the illuminated area. Activating the third condenser (C3) adds a lens between C2 and the objective. The illuminated-area slider then sets the focal lengths of C2 and C3 such that the illumination at the specimen remains parallel while the beam diameter changes, which is the condenser zoom described above. The sliders of the two lenses can still be changed individually, which shows how the illumination becomes convergent or divergent when the zoom condition is not fulfilled. With only two condensers, a change of the beam diameter is always accompanied by a change of the convergence.
Interactive element
Click the rocket symbol at the top of the page and select Live Code, or expand the Show code toggle and click ▶ to run the cell. Adjust the sliders to see how condenser focal lengths and apertures shape the electron beam, and activate Third condenser (C3) to try the condenser zoom.
8.3. Projector system and detection#
Below the objective, a series of intermediate and projector lenses magnify and relay the image to the detector at the bottom of the column (Section 8.3.1). Total magnifications of 10,000–500,000× are routinely available. Many cryo-TEM instruments also contain an energy filter in front of the detector, which removes inelastically scattered electrons with a slit (typically 10–20 eV wide). Because these electrons have lost energy in the specimen, they are focused differently by the chromatic lens and contribute a blurred background, and removing them improves the contrast, in particular of thick specimens.
8.3.1. Electron detectors#
The image is recorded by an electronic detector in the image plane of the projector system. Two properties describe how well it preserves the information in the incident electrons. The modulation transfer function (MTF) is the transfer function of the detector, the Fourier transform of its point spread function (Section 9.4), and it states how strongly the contrast of each spatial frequency \(\nu\) is reduced. The detective quantum efficiency (DQE) states how much of the signal-to-noise ratio (SNR, the ratio of signal power to noise power, Section 7.2) of the incident electrons is retained,
A perfect detector has DQE = 1 at all frequencies. Because the dose in cryo-EM is low (Section 7.1), every electron carries information and a high DQE, in particular up to the Nyquist frequency (Section 9.2), directly reduces the number of particles that is needed for a given resolution. Fig. 8.3 compares typical DQE curves of the detectors that are described below.
Fig. 8.3 Schematic DQE shown as a function of spatial frequency, in units of the physical Nyquist frequency of the detector. The curves show typical trends for different detector types. The DQE of a real detector depends on the accelerating voltage and the dose rate. In super-resolution mode, counting detectors can transfer frequencies above the physical Nyquist frequency.#
Photographic film was the standard for cryo-EM for a long time. It offers a high resolution, but it has to be developed and digitised, and it cannot record movies. Scintillator-based cameras (CCD or CMOS) convert each electron into light in a scintillator, and this light spreads on its way to the sensor. The resulting blurring lowers the DQE at high spatial frequencies. Direct electron detectors (DEDs) detect the electrons in a thinned silicon sensor without this conversion. The signal of each electron is confined to a few pixels, which gives a DQE that is considerably higher than for scintillator-based cameras. Their introduction around 2012 was an important reason for the rapid improvement in resolution of cryo-EM. Fig. 8.4 compares the two designs.
Fig. 8.4 Indirect and direct electron detectors. In the indirect detector, the electron generates light in a scintillator, which spreads on its way through the fibre-optic plate to the sensor. In the direct detector, the electron is detected in the sensor itself, so that its signal is confined to one or a few pixels.#
DEDs can be operated in two modes. In integrating mode, the charge that is collected in each pixel is summed, and the signal of an electron fluctuates because electrons deposit different amounts of energy. In counting mode, the position of each electron is determined and each electron is counted once, independent of the signal it produced, which removes this fluctuation and increases the DQE. Counting requires a dose rate that is low enough that electrons rarely arrive in the same pixel during one frame (coincidence loss), and the electron position can be determined with sub-pixel precision, which makes it possible to record at an effectively smaller pixel size (super-resolution mode). The steps of Fig. 8.5 illustrate how the signal of a single electron is processed in the two modes.
Fig. 8.5 From the electron to the pixel values. (A) An electron hits the sensor. (B) Its signal spreads over several pixels. © In integrating mode, the pixel values (here normalised to a sum of one) are recorded, so that the signal depends on the energy that the electron deposited. (D) In counting mode, the centroid of the cluster is determined. (E) The electron is counted once, at the position of the centroid; in super-resolution mode, this position is assigned to a sub-pixel.#
Modern detectors are fast enough to record the exposure as a movie of many short frames (dose fractionation). The specimen moves during the exposure, because of stage drift and beam-induced motion, and the frames can be aligned before they are summed, so that this motion does not blur the high-resolution information. The frames can also be weighted according to the accumulated dose, as the high-resolution information is damaged first (Section 7.1).
The magnification that is chosen determines the pixel size at the specimen, which is the pixel size of the detector divided by the magnification. A detector with pixels of 5 µm at an effective magnification of 100,000× yields 0.5 Å per pixel. To record a resolution \(d\), the pixel size must not exceed \(d/2\) (Section 9.2), but a smaller pixel size also reduces the field of view, so that the magnification is a trade-off between the target resolution and the number of particles per image.